6. PARAMETER DESCRIPTION

(3) Display and output condition for setting alarm occurrences

 

 

Alarm description

Selection

Alarm

4-20 mA

Totali-

Pulse

Display

Alarm

 

 

(parameter

output

output

zation

output

unit

record

 

 

 

number)

 

 

 

 

 

 

50

Span > 10m/s

Span flow velocity setting

 

 

 

 

 

 

 

is 11 m/s or more

 

Closed

Fixed

Stopped

Stopped

Alarm Mode

No

 

 

 

 

 

 

Span flow velocity setting

 

51

Span < 0.1m/s

 

(On)

 

 

 

(message)

 

is 0.05 m/s or less

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

52

TTL>10000p/s

Totalization rate is 11000

 

 

 

 

 

 

 

pps or more

 

Closed

Normal

 

Normal

Alarm Mode

 

 

 

 

Stopped

No

 

 

 

 

53

TTL<0.0001p/s

Totalization rate is

 

(On)

operation

operation

(message)

0.00005 pps or less

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

54

4-20 Lmt Err

The condition [4-20 low

 

Closed

Fixed

Normal

Normal

Alarm Mode

No

limit (J11) < 4-20 high

 

operation

operation

(message)

 

 

limit (J12)] is not satisfied.

 

(On)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The condition [No. 1 range

 

 

 

 

 

 

 

55

Multi Rng Err

< No. 2 range < No. 3 range

 

Closed

Fixed

Stopped

Stopped

Alarm Mode

No

< No. 4 range] is not

 

 

 

satisfied for multiple ranges.

 

(On)

 

 

 

(message)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The condition [High Alarm

 

 

 

 

 

 

 

 

 

(G11) – Low Alarm (G10) >

 

 

 

 

 

 

 

 

 

H/L Alarm Hys (G14)] or the

 

Closed

Normal

Normal

Normal

Alarm Mode

 

56

H/L HH/LL Set

condition [High High Alarm

 

No

(G13) – Low Low Alarm

 

(On)

operation

operation

operation

(message)

 

 

 

 

 

 

(G12) > H/L Alarm Hys

 

 

 

 

 

 

 

 

 

(G14)] is not satisfied.

 

 

 

 

 

 

 

 

 

Mass units have been

 

Closed

 

 

 

Alarm Mode

 

57

Dens Set Err

selected for Base Flow Unit

 

Fixed

Stopped

Stopped

No

 

(On)

(message)

 

 

(C40) but density is set to 0.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulse rate is 11000 pps

 

 

 

 

 

 

 

 

 

or more with 50% duty

 

 

 

 

 

 

 

60

PLS >10000p/s

selection.

NO

 

 

 

 

 

 

Pulse rate is 10000 pps

(G30)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

or more with 0.05 ms

 

 

 

 

 

 

 

 

 

selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

61

PLS > 5000p/s

Pulse rate is 5000 pps or

 

 

 

 

 

 

 

more with 0.1 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

62

PLS > 1000p/s

Pulse rate is 1000 pps or

 

 

 

 

 

 

 

more with 0.5 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Closed

Normal

Normal

 

Alarm Mode

 

63

PLS > 500p/s

Pulse rate is 500 pps or

 

Stopped

No

 

(On)

operation

operation

(message)

more with 1 ms selection.

 

 

 

 

 

 

 

 

 

 

 

64

PLS > 25p/s

Pulse rate is 25 pps or

 

 

 

 

 

 

 

more with 20 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

65

PLS > 15p/s

Pulse rate is 15 pps or

 

 

 

 

 

 

 

more with 33 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

66

PLS > 10p/s

Pulse rate is 10 pps or

 

 

 

 

 

 

 

more with 50 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

67

PLS > 5p/s

Pulse rate is 5 pps or

 

 

 

 

 

 

 

more with 100 ms selection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

70

PLS<0.0001p/s

Pulse rate is 0.00005 pps

 

 

 

 

 

 

 

or less.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure Mode (C20) is

 

 

 

 

 

 

 

71

Meas Mod Set

set to Enhanced DF

 

Closed

Fixed

Stopped

Stopped

Alarm Mode

No

without selecting an

 

 

(On)

(message)

 

 

optional code HF1 or

 

 

 

 

 

 

 

HF2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

72

Size Set Err

A value of 3000.1 mm or

 

Closed

Fixed

Stopped

Stopped

Alarm Mode

No

 

 

more is set for Nominal

 

(On)

 

 

 

(message)

 

 

 

Size (C32).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The condition [Level:1<

 

Closed

Normal

Normal

Normal

Alarm Mode

 

73

Adh Set Err

Level:2<Level:3<Level:4]

 

No

is not satisfied for

 

(On)

operation

operation

operation

(message)

 

 

 

 

 

 

adhesion diagnostic level.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Occurring of

YES

Open

Fixed

Stopped

Stopped

Alarm Mode

No

 

any alarm from

 

(G30)

(Off)

(message)

 

50 through 73

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm Out Act Mode.

4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.

T0650.EPS

6-37

IM 01E20C01-01E

Page 65
Image 65
APC AXFA11G Display and output condition for setting alarm occurrences, Alarm description Selection, Output Unit