8.3.4.4Example 1 of Other Settings: Meter Factor

8. OPERATION VIA HART COMMUNICATOR

8.3.4.4Example 1 of Other Settings: Meter Factor

The “Meter Factor” is engraved on the data plate of the combined flowtube.

The meter factor is required to compute the correct electromotive force proportional to the fluid velocity, and is determined by actual flow-test calibration at the factory.

Open the “MF” setting display:

1. Device Setup Select “ Standard DF ”.Manual backgroundManual background 4. Detailed Setup Manual backgroundManual backgroundManual background 1. Basic Setup Manual backgroundManual backgroundManual background 3. MF Set

1

AXFA11 : FI-1201

Basic Setup

1Tag

2PV Damping

3Manual backgroundMF Set

4Select Flow Tube

5 Nominal Size Unit

SAVE HOME

or

YZ/

3

Select “ MF Set ”.

2AXFA11 : FI-1201 Select Measure Mode

Standard DF Enhanced DF

ABORT ENTER

3AXFA11 : FI-1201 Input Low MF value

1.0000

1.1111

DEL ABORT ENTER

4AXFA11 : FI-1201 Input High MF value

1.0000

1.2222

DEL ABORT ENTER

5AXFA11 : FI-1201

Basic Setup

1Tag

2PV Damping

3Manual backgroundMF Set

4Select Flow Tube

Manual background5 Nominal Size Unit

SAVE HOME

STU or F4 1

(ENTER)

F4

(ENTER)

F4

(ENTER)

F3

(HOME)

Select “ Standard DF ”.

Enter “1.1111” with alphanumeric keys. Press ENTER (F4).

Enter “1.2222” with alphanumeric keys. Press ENTER (F4).

Press HOME (F3), and return to the “Online” menu.

F0812.EPS