Adhesion Diagnostic Function

Setting

Function

 

 

No

Halt the adhesion diagnostic function

 

 

Yes

Carry out the adhesion diagnostic function

 

 

 

T0642.EPS

NOTE

Adhesion Diagnostic Function

￿This function diagnose adhesion using elec- trode resistance values.

￿When “Adhesion check” has been set forB41/ H11: Display Select 2 or B42/H12: Display Select 3, the diagnose adhesion is indicated on the display unit using four different levels.

￿If the judgment value for Level 3 is exceeded, a warning is displayed; and if the value for Level 4 is exceeded, an alarm is displayed.

￿Available conductivity for this function is limited to:

Nominal size 10 mm or smaller: 30 S/cm

Nominal size 15 mm or larger: 10 S/cm Make sure to use the adhesion diagnostic function with the greater conductivity than the above mentioned value.

 

 

 

 

Warning

Alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Level: 1

Level: 2

Level: 3

Level: 4

F0619.EPS

[K11: Adhesion Level1] Setting the resistance value for adhesion level 1.

This parameter sets the resistance value (in M ohm) for judgment of Level 1.

[K12: Adhesion Level2] Setting the resistnce value for adhesion diagnostic level 2.

This parameter sets the resistance value (in M ohm) for judgment of Level 2.

[K13: Adhesion Level3] Setting the resistance value for adhesion diagnostic level 3.

This parameter sets the resistance value (in M ohm) for judgment of Level 3.

*: The warning 80: Adhesion Wng is displayed when the adhension level reaches Level 3.

*: If “Warning Output” has been selected for F10:

SO1 Function or F11: SO2 Function, then

status output will be performed when the adhesion level reaches Level 3.

6. PARAMETER DESCRIPTION

[K14: Adhesion Level4] Setting the resistance value for adhesion diagnostic level 4.

This parameter sets the resistance value (in M ohm) for judgment of Level 4.

*: The process alarm 33: Adhesion Alm is displayed when the adhension level reaches Level 4.

*: Alarm output will be performed if “Yes” has been selected for G34: Alm-Adhesion.

[K15: Adh Measure Value] Displays the resistance value for adhesion diagnose

This parameter displays the value measured using the adhesion diagnostic function (in M ohm).

When “No” is selected for K10: Adhesion Check, this parameter displays the indetermination value.

(10)Menu M: Automatic Zero Adjustment Function Setting items

Menu M contains items that are relevant to automatic adjustment of the zero point.

[M10: Auto Zero Exe] Execution of automatic zero adjustment function

Refer to the description of parameter B50.

[M11: Magflow Zero] Results of automatic zero adjustment

This parameter is used to display the results obtained from B50/M10: Auto Zero Exe. Specifically, the correction value displayed, and it is also possible to directly enter correction value.

(11) Menu N: Loop Test Setting items

Menu N contains items that are relevant to loop testing.

[N10: Test Mode] Setting for loop test execution

Setting

Function

 

 

Normal

No execution of loop testing

 

 

Test

Loop testing is started

 

 

T0644.EPS

IMPORTANT

(1)Test output has priority over flow rate mea- surement signals. When carrying out flow rate measurements, be sure to always return to “Normal”.

(2)Upon entry to the Test Mode, all output terminals will simultaneously adopt test condition.

6-33

IM 01E20C01-01E