APC AXFA11G user manual “Returning fld dev to original output” appears, ‘19.050’, 8-12

Models: AXFA11G

1 118
Download 118 pages 18.28 Kb
Page 86
Image 86
‘19.050’

6AXFA11 : FI-1201

Fld dev output 4.000 mA equal to reference meter?

1 Yes

2 No

ABORT ENTER

7AXFA11 : FI-1201 Setting fld dev output to 20mA

ABORT OK

8AXFA11 : FI-1201 Enter meter value 20.000

HELP DEL ABORT ENTER

9AXFA11 : FI-1201

Fld dev output 20.000 mA equal to reference meter?

1 Yes

2 No

ABORT ENTER

10AXFA11 : FI-1201 NOTE-Loop may be returned to automatic control

OK

11AXFA11 : FI-1201

Adjustment

1 Auto Zero Exe

2 Magflow Zero

3“Returning fld dev to original output” appears.D/A trim

4 Scaled D/A trim

HELP SAVE HOME

F4

(ENTER)

F4

(OK)

‘19.050’

F4

(ENTER)

F4

(ENTER)

F4

(OK)

F3

(HOME)

8. OPERATION VIA HART COMMUNICATOR

Ammeter reading: 4.000

Because the reading on the ammeter is 4.000 mA, select “Yes” and press ENTER (F4).

If the reading is not 4.000 mA, select “No”. Repeat step 5 until the ammeter reads 4.000 mA.

Press OK (F4), and the AXFA11 outputs an output signal of 100%.

Ammeter reading: 19.050

Carry out the same procedures as those described in 5 .

Ammeter reading: 20.000

Returning fld dev to original output” appears.

Press OK (F4).

Press HOME (F3), and return to the “Online” menu.

F0814-3.EPS

8-12

IM 01E20C01-01E