Liebert Universal Monitor Set Up Alarmable Inputs in Latched or Unlatched Mode, Set Up Delay Time

Models: Universal Monitor

1 144
Download 144 pages 39.86 Kb
Page 55
Image 55

System and Control Options

9.3.4 Set Up Alarmable Inputs in Latched or Unlatched Mode

Setup Input Menu

 

Alarmable input points may be set up in Latched mode (Y), which requires the user to clear the Universal Monitor alarms after an alarm has occurred, or Unlatched mode (N), in which alarms will automati- cally clear after a return-to-normal state. The default setting for alarmable inputs is N (Unlatched). Any input defined as an event is automatically Unlatched.

A digital input defined as Latched does not reset automatically after an alarm. It will remain in the triggered position even after the moni- tored device returns to normal. After an alarm, the user must manu- ally reset Latched inputs, as described in 9.12.1 - Clear Active Alarms. This also means that any digital outputs linked to a Latched digital input will remain triggered until the input is manually reset.

A digital input defined as Unlatched will automatically track the sta- tus of the monitored device. No action by the user is required.

From the Setup Input Menu, choose Latched, as shown above right, and press Enter .

To change this feature, use the arrows ↑ ↓ to choose Y (Yes - Latched) or N (No - Unlatched), then press Enter .

NOTE: Only alarmable input points may be set up as Latched. If you attempt to set up an event as Latched, the message shown at right appears. Press the ESC key to return to the previous screen to con- tinue configuring the digital input.

The only way to reset latched alarms is by clearing active alarms. This step is required to return latched points to their normal state. For details, see 9.12.1 - Clear Active Alarms.

9.3.5Set Up Delay Time

SETUP INPUT_01

LABEL: INPUT_01

NORM OPEN/CLOSE: NO

ALARM/EVENT: AL

↑ ↓=NEXT

=SELECT

>LATCHED: N

DELAY: 00:01

Latched / Unlatched

IS ALARM

LATCHED? N

↑ ↓=Y/N

=END

Message (if input has been defined as an event)

LATCH NOT ALLOWED WHEN IN EVENT MODE

ESC=BACK

Setup Input Menu

Each input may be set up with a specified time delay between the occurrence of an alarm or event—a monitored device changes state— and any response by the Universal Monitor. This feature can prevent needless responses to transient conditions.

This delay is the amount of time that must elapse before the Universal Monitor acknowledges the change of state in the monitored device and takes any further action, such as dialing configured pagers. The default time delay is 1 second (displayed as 00 MIN : 01 SEC).

The time delay applies only when a monitored device changes from its normal state—for example, the Universal Monitor detects a closed contact that is defined as Normally Open.

When the monitored device returns to its normal state, the Universal Monitor acknowledges the change of state immediately.

From the Setup Input Menu, choose Delay, as shown at right, and press Enter .

SETUP INPUT_01

LABEL: INPUT_01

NORM OPEN/CLOSE: NO

ALARM/EVENT: AL

↑ ↓=NEXT

=SELECT

LATCHED: N

DELAY: 00:01

Delay Time

DELAY TIME BEFORE TRIGGER?

00 MIN : 01 SEC

=SELECT =NEXT/END

To change the time delay in minutes and seconds (in the range of 00:00 to 99:59):

For minutes (MIN), press and hold one of the arrow keys ↑ ↓ scroll through the numbers 00-99.

Press Enter to advance to seconds (SEC).

For seconds (SEC), press and hold one of the arrow keys ↑ ↓ scroll through the numbers 00-59.

When finished, press Enter again.

47

Page 55
Image 55
Liebert Universal Monitor Set Up Alarmable Inputs in Latched or Unlatched Mode, Set Up Delay Time, Latched / Unlatched