Abdomen

AbdomenSeq

Indications:

Sequential mode for all routine studies in the region of abdomen, e.g., evaluation, follow-up examinations etc.

The whole scan length covers 15.2 cm.

 

AbdSeq

 

 

kV

130

 

 

Effective mAs

70

 

 

Rotation time

1.0 sec

 

 

Slice collimation

4.0 mm

 

 

Slice width

8.0 mm

 

 

Feed/Scan

8.0 mm

 

 

Kernel

B41s

 

 

CTDIVol

7.55 mGy

 

 

Effective dose

Male: 2.35 mSv

 

Female: 2.75 mSv

 

 

Contrast medium IV injection

 

 

Start delay

50 – 60 sec.

 

 

Flow rate

3.0 ml/sec.

 

 

Total amount

100 ml

 

 