It is mandatory to position the patient head in the center of the scan field – use the lateral laser light marker for positioning.

Gantry tilt is not necessary since you have the ability to tilt the reference line to generate an axial refor- matted image at the desired plane. However, in order to minimize the scan length for the same ana- tomical region, it is recommended to position the patient’s head at the appropriate scan plane when- ever possible:

– For the upper and lower jaw:

occlusal plane in parallel to the scan plane.

– For either jaw:

jaw bone in parallel to the scan plane.

It is recommended to end the exam first, and then start the Dental evaluation.

User interface of syngo Dental CT