Lower Extremities

ExtrRoutineHR

Indications:

Spiral mode for HiRes bone study, e.g., trauma, ortho- pedic indications etc.

A range of 10 cm will be covered in 52 sec.

 

ExtrHR

2nd reconstruction

kV

130

 

 

 

 

Effective mAs

30

 

 

 

 

Rotation time

1.0 sec

 

 

 

 

Slice collimation

1.0 mm

 

 

 

 

Slice width

2.0 mm

1.25 mm

 

 

 

Feed/Rotation

2.0 mm

 

 

 

 

Pitch Factor

1.0

 

 

 

 

Increment

2.0 mm

0.8 mm

 

 

 

Kernel

U90s

U90s

 

 

 

CTDIVol

3.23 mGy

 

 

 

Effective dose

Male: 0.01 mSv

 

Female: 0.01 mSv

Hint

For image reconstruction of soft tissue, use kernel B31s.

161

Page 161
Image 161
Siemens Software version syngo CT 2005C manual For image reconstruction of soft tissue, use kernel B31s 161