Children

Hints

Please change the mAs value according to the weight of the child:

35 – 44 kg 45 mAs

45 – 54 kg 60 mAs.

Children with a body weight of more than 55 kg should be examined with an adult protocol.

CARE Bolus may be used to optimize the bolus tim- ing. Set the ROI for the monitoring scan in the abdominal aorta with triggering threshold of 120 HU, or use manual triggering.

235

Page 235
Image 235
Siemens Software version syngo CT 2005C manual Children