Siemens Software version syngo CT 2005C manual HeadRoutineBaby

Models: Software version syngo CT 2005C

1 321
Download 321 pages 19.7 Kb
Page 192
Image 192

Children

HeadRoutine_Baby

Indications:

Routine head spiral studies, e.g., brain tumors, cranial trauma, hydrocephalus, hemorrhaging and other abnormalities etc.

A range of 8 cm will be covered in 27 sec.

 

Head

 

 

kV

80

 

 

Effective mAs

122

 

 

Rotation time

1.5 sec

 

 

Slice collimation

2.5 mm

 

 

Slice width

3.0 mm

 

 

Feed/Rotation

5.0 mm

 

 

Pitch Factor

1.0

 

 

Increment

3.0 mm

 

 

Kernel

C30s

 

 

CTDIVol

7.89 mGy

 

 

Effective dose

Male: 0.84 mSv*

 

Female: 0.92 mSv*

 

 

*The conversion factor for an 8-week-old child, and a scan range of 80 mm was used.

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/

 

Access site

 

 

Total amount

1 – 2 ml per kg of body weight

192

Page 192
Image 192
Siemens Software version syngo CT 2005C manual HeadRoutineBaby