Children

HeadSeq_Child

Indications:

Sequential mode for routine head studies for children, e.g., tumors, hydrocephalus, hemorrhaging, abnor- malities, etc.

A range of 8.25 cm is predefined.

 

HeadSeq

 

 

kV

130

 

 

Effective mAs

120

 

 

Rotation time

1.5 sec

 

 

Slice collimation

2.5 mm

 

 

Slice width

2.5 mm

 

 

Feed/Scan

5.0 mm

 

 

Kernel

C30s

 

 

CTDIVol

27.32 mGy

 

 

Effective dose

Male: 1.09 mSv*

 

Female: 1.18 mSv*

*The conversion factor for a 7-year-old child, and a scan range of 82.5 mm was used.

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/

 

Access site

 

 

Total amount

1 – 2 ml per kg of body weight

198

Page 198