Spine

 

C5-6

C6-7

 

 

 

kV

130

130

 

 

 

Effective mAs

170

190

 

 

 

Rotation time

1.5 sec

1.5 sec

 

 

 

Slice collimation

1.0 mm

1.0 mm

 

 

 

Slice width

2.0 mm

2.0 mm

 

 

 

Feed/Scan

2.0 mm

2.0 mm

 

 

 

Kernel

B50s

B50s

 

 

 

CTDIVol

18.33 mGy

20.48 mGy

 

 

 

Effective dose

Male:

Male:

 

0.66 mSv

0.30 mSv

 

Female:

Female:

 

0.73 mSv

0.50 mSv

 

 

 

Hints

Dental artifacts can be reduced by tilting the gantry.

For MPRs and 3D post-processing, use spiral mode with thinner slice and reconstruct images with 50% overlap in image reconstruction. Define a long range to cover the levels of interest in one block.