Vascular

ThorAngio/ThorAngio08s

Indications:

Spiral study for location and extent of aneurysms, dis- section, and ruptures of the thoriac aorta.

A range of 20 cm will be covered in 24/19 sec.

 

ThorAngio

2nd reconstruction

kV

130

 

 

 

 

Effective mAs

55

 

 

 

 

Rotation time

0.8/1.0 sec

 

 

 

 

Slice collimation

2.5 mm

 

 

 

 

Slice width

5.0 mm

3.0 mm

 

 

 

Feed/Rotation

9.0 mm

 

 

 

 

Pitch Factor

1.8

 

 

 

 

Increment

5.0 mm

2.0 mm

 

 

 

Kernel

B31s

B31s

 

 

 

CTDIVol

5.93 mGy

 

 

 

Effective dose

Male: 1.91 mSv

 

Female: 2.67 mSv

 

 

 