Children

NeonateBody

Indications:

Spiral mode for routine neonate body studies, e.g., tumors, abnormalities, malformations, abscesses, etc.

A range of 15 cm will be covered in 17 sec.

 

NeonateBody

kV

80

 

 

Effective mAs

41

 

 

Rotation time

1.0 sec

 

 

Slice collimation

2.5 mm

 

 

Slice width

5.0 mm

 

 

Feed/Rotation

10.0 mm

 

 

Pitch Factor

2.0

 

 

Increment

5.0 mm

 

 

Kernel

B41s

 

 

CTDIVol

1.13

 

 

Effective dose

Male: 1.59 mSv*

 

Female: 2.01 mSv*

*The conversion factor for a 7-year-old child, and a scan range of 150 mm was used.

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/

 

Access site

 

 

Total amount

1 – 2 ml per kg of body weight

238