Head

InnerEarHR

Indications:

Spiral mode for inner ear High Resolution studies, e.g., inflammatory changes, tumorous processes of pyra- mids, cerebellopontine angle tumors, post-traumatic changes, etc.

A range of 2.5 cm will be covered in 22 sec.

 

InnerEar

2nd reconstr.

kV

130

 

 

 

 

Effective mAs

70

 

 

 

 

Rotation time

1.5 sec

 

 

 

 

Slice collimation

1.0 mm

 

 

 

 

Slice width

2.0 mm

1.25 mm

 

 

 

Feed/Rotation

2.0 mm

 

 

 

 

Pitch Factor

1.0

 

 

 

 

Increment

2.0 mm

0.8 mm

 

 

 

Kernel

H80s

H80s

 

 

 

CTDIVol

15.94 mGy

 

 

 

Effective dose

Male: 0.17 mSv

 

Female: 0.20 mSv

76

Page 76
Image 76
Siemens Software version syngo CT 2005C manual InnerEarHR, InnerEar 2nd reconstr