Specials

A range for the Thorax of 25 cm will be covered in 15 sec., a range for the AbdPelvis of 20 cm will be cov- ered in 12 sec.

 

Thorax

AbdPelvis

 

 

 

kV

130

130

 

 

 

Effective mAs

40

60

 

 

 

Rotation time

1.0 sec

1.0 sec

 

 

 

Slice collimation

5.0 mm

5.0 mm

 

 

 

Slice width

8.0 mm

8.0 mm

 

 

 

Feed/Rotation

20.0 mm

20.0 mm

 

 

 

Pitch Factor

2.0

2.0

 

 

 

Increment

8.0 mm

8.0 mm

 

 

 

Kernel

B41s

B41s

 

 

 

CTDIVol

4.31 mGy

6.47 mGy

 

 

 

Effective dose

Male:

Male:

 

1.82 mSv

2.59 mSv

 

Female:

Female:

 

2.45 mSv

3.31 mSv

For long range scanning, please pay attention to the mark of scannable range on the table mattress while positioning the patient.

In some cases, it might be advisable to position the patient feet first so that there will be more space for the intensive care equipment around.

179

Page 179
Image 179
Siemens Software version syngo CT 2005C manual Thorax AbdPelvis