Children

InnerEar

Indications:

High Resolution Spiral mode for inner ear studies, e.g., malformations of the inner ear, inflammatory changes, pathologies of the mastoid process, tumor processes of the pyramids, post-traumatic changes, etc.

A range of 2.5 cm will be covered in 22 sec.

 

InnerEar

 

 

kV

80

 

 

Effective mAs

125

 

 

Rotation time

1.5 sec

 

 

Slice collimation

1.0 mm

 

 

Slice width

1.25 mm

 

 

Feed/Rotation

2.0 mm

 

 

Pitch Factor

1.0

 

 

Increment

1.0 mm

 

 

Kernel

H80s

 

 

CTDIVol

8.09 mGy

 

 

Effective dose

Male: 0.04 mSv*

 

Female: 0.04 mSv*

 

 

*The conversion factor for a 7-year-old child, and a scan range of 25 mm was used.

200

Page 200
Image 200
Siemens Software version syngo CT 2005C manual InnerEar