Siemens Software version syngo CT 2005C manual Body Kernels

Models: Software version syngo CT 2005C

1 321
Download 321 pages 19.7 Kb
Page 191
Image 191

Children

Body Kernels

As standard kernels for body tissue studies B31s or B41s is recommended; softer images are obtained with B20s.

For higher sharpness, as is required e.g., in patient protocols for cervical spine, shoulder, extremities, thorax, the kernels B50s, B60s, B70s are available.

The special kernels are mostly used for ”physical” mea- surements with phantoms, e.g., for adjustment proce- dures (S80s), for constancy and acceptance tests (S80s, U90s), or for specification purposes (U90s).