Children

BodyAngio

Indications:

Spiral mode for abdominal CT Angio studies, e.g., vas- cular abnormalities, aneurysms, etc.

A range of 10 cm will be covered in 19 sec.

 

BodyAngio

2nd reconstr.

kV

80

 

 

 

 

Effective mAs

41

 

 

 

 

Rotation time

1.0 sec

 

 

 

 

Slice

1.5 mm

 

collimation

 

 

 

 

 

Slice width

3.0 mm

2.0 mm

 

 

 

Feed/Rotation

6.0 mm

 

 

 

 

Pitch Factor

2.0

 

 

 

 

Increment

3.0 mm

1.5 mm

 

 

 

Kernel

B41s

B41s

 

 

 

CTDIVol

1.13 mGy

 

 

 

 

Effective dose

Male:

 

0.73mSv*

Female:

0.84mSv*

* The conversion factor for a 7-year-old child, and a scan range of 100 mm was used.

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/

 

Access site

 

 

Total amount

1 – 2 ml per kg of body weight

234

Page 234
Image 234
Siemens Software version syngo CT 2005C manual BodyAngio 2nd reconstr