Pelvis

SI_Joints

Indications:

Spiral mode for bone studies of the sacroiliac joints.

A typical range of 8 cm will be covered in 18 sec.

 

SI_Joints

2nd reconstruction

kV

130

 

 

 

 

Effective mAs

80

 

 

 

 

Rotation time

1.0 sec

 

 

 

 

Slice

2.5 mm

 

collimation

 

 

 

 

 

Slice width

3.0 mm

3.0 mm

 

 

 

Feed/Rotation

5.0 mm

 

 

 

 

Pitch Factor

1.0

 

 

 

 

Increment

3.0 mm

2.0 mm

 

 

 

Kernel

B70s

B70s

 

 

 

CTDIVol

8.62 mGy

 

 

 

 

Effective dose

Male:

 

 

1.44 mSv

 

 

Female:

 

 

1.93 mSv

 

 

 

 

130

Page 130
Image 130