Children

ExtrHR_Baby

Indications:

Spiral mode for high resolution extremity studies, e.g., post-traumatic changes, tumors, malformations, etc.

A range of 10 cm will be covered in 27 sec.

 

ExtrHR

kV

80

 

 

Effective mAs

50

 

 

Rotation time

1.0 sec

 

 

Slice collimation

1.0 mm

 

 

Slice width

1.25 mm

 

 

Feed/Rotation

4.0 mm

 

 

Pitch Factor

2.0

 

 

Increment

0.8 mm

 

 

Kernel

B70s

 

 

CTDIVol

1.38 mGy

 

 

Effective dose

Male: < 0.01 mSv*

 

Female: < 0.01 mSv*

*The conversion factor for an 8-week-old child, and a scan range of the knee of 100 mm was used.

222

Page 222
Image 222
Siemens Software version syngo CT 2005C manual ExtrHRBaby