Siemens Software version syngo CT 2005C manual ThorRoutine, Reconstruction

Models: Software version syngo CT 2005C

1 321
Download 321 pages 19.7 Kb
Page 103
Image 103

Thorax

 

ThorRoutine

2nd

 

 

reconstruction

 

 

 

kV

130

 

 

 

 

Effective mAs

60

 

 

 

 

Rotation time

0.8/1.0 sec

 

 

 

 

Slice

4.0 mm

 

collimation

 

 

 

 

 

Slice width

5.0 mm

5.0 mm

 

 

 

Feed/Rotation

14.4 mm

 

 

 

 

Pitch Factor

1.8

 

 

 

 

Increment

5.0 mm

5.0 mm

 

 

 

Kernel

B41s

B70s

 

 

 

CTDIVol

6.47 mGy

 

 

 

 

Effective dose

Male:

 

 

3.28 mSv

 

Female: 4.23 mSv

Contrast medium IV injection

Start delay

25 – 30 sec.

Flow rate

2.5 ml/sec.

 

 

Total amount

80 ml

Hints

You could repeat the same protocol by simply click- ing the chronicle with the right mouse button for “repeat“, e.g., when both non-contrast and contrast studies are required.

For lung cancer evaluation, this protocol can be com- bined with the protocol “Neck Routine”.