Head

Dental

This is the scan protocol for the syngo Dental CT appli- cation package. It is used for evaluation and reformat- ting of the upper and lower jaws.

It enables the display and measurement of the bone structures of the upper and lower jaw as the basis for planning in oral surgery.

Indications:

Spiral mode for dental studies.

A range of 5 cm will be covered in 27 sec.

 

Dental

 

 

kV

130

 

 

Effective mAs

45

 

 

Rotation time

1.0 sec

 

 

Slice collimation

1.0 mm

 

 

Slice width

1.25 mm

 

 

Feed/Rotation

2.0 mm

 

 

Pitch Factor

1.0

 

 

Increment

0.5 mm

 

 

Kernel

H60s

 

 

CTDIVol

10.25 mGy

 

 

Effective dose

Male: 0.13 mSv

 

Female: 0.16 mSv

 

 

Load the study into the application “syngo Dental CT”. For further information, please refer to the chapter "syngo Dental CT".