Head

 

Base

Cerebrum

 

 

 

kV

130

130

 

 

 

Effective mAs

110

110

 

 

 

Rotation time

1.5 sec

1.5 sec

 

 

 

Slice collimation

1.5 mm

2.5 mm

 

 

 

Slice width

3.0 mm

8.0 mm

 

 

 

Feed/Rotation

3.0 mm

5.0 mm

 

 

 

Pitch Factor

1.0

1.0

 

 

 

Increment

3.0 mm

8.0 mm

 

 

 

Kernel

H31s

H31s

 

 

 

CTDIVol

25.05 mGy

25.05 mGy

 

 

 

Effective dose

Male:

Male:

 

0.37 mSv

0.70 mSv

 

Female:

Female:

 

0.38 mSv

0.77 mSv

 

 

Contrast medium IV injection

 

 

 

 

Volume

50 – 60 ml

 

 

 

 

Flow rate

2 ml/sec.

 

 

 

 

Start delay

50 – 60 sec.

 

 

 

 

Hints

An automatic bone correction allows for improved head image quality, without any additional post-pro- cessing.

In order to optimize image quality versus radiation dose, scans are provided within a maximum scan field of 300 mm with respect to the iso-center. No recon job with a field of view exceeding those limits will be possible. Therefore, patient positioning has to be performed accurately to ensure a centered loca- tion of the skull.