Siemens Software version syngo CT 2005C manual Contrast medium IV injection

Models: Software version syngo CT 2005C

1 321
Download 321 pages 19.7 Kb
Page 165
Image 165

Vascular

Contrast medium IV injection

Start delay

18 sec.

Flow rate

3.5 ml/sec.

 

 

Total amount

75 ml

Hints

Use of CARE Bolus with monitoring scans positioned at the level of the basilar artery or carotid artery. Set the trigger threshold at 120 HU, or use manual trig- gering.

An automatic bone correction and an advanced algo- rithm allow for improved head image quality, with- out any additional post-processing.

165

Page 165
Image 165
Siemens Software version syngo CT 2005C manual Contrast medium IV injection