Abdomen

 

Non

Arterial

Venous

 

Contrast

Phase

Phase

 

 

 

 

 

kV

130

130

130

 

 

 

 

 

 

Effective mAs

80

80

80

 

 

 

 

 

 

Rotation time

1.0 sec

0.8/1.0

1.0 sec

 

 

sec

 

 

 

 

 

 

 

Slice

5.0 mm

5.0 mm

5.0 mm

collimation

 

 

 

 

 

 

 

 

 

Slice width

8.0 mm

6.0 mm

6.0 mm

 

 

 

 

 

Feed/Rotation

15.0 mm

14.0/

15.0 mm

 

 

15.0 mm

 

 

 

 

 

 

 

Pitch Factor

1.5

1.4/1.5

1.5

 

 

 

 

 

 

Increment

8.0 mm

6.0 mm

6.0 mm

 

 

 

 

 

Kernel

B41s

B41s

B41s

 

 

 

 

 

CTDIVol

8.62 mGy

8.62 mGy

8.62 mGy

 

 

 

 

 

Effective dose

Male:

Male:

Male:

 

3.34 mSv

3.34 mSv

3.34 mSv

 

Female:

Female:

Female:

 

4.19 mSv

4.19 mSv

4.19 mSv

 

 

 

Contrast medium IV injection

 

 

 

 

 

Start delay

20 – 25 sec.* (arterial phase)

 

50 – 75 sec.* (venous phase)

 

 

 

 

Flow rate

3.0 ml/sec.

 

 

 

 

 

 

 

Total amount

100 ml

 

 

 

*CARE Bolus may be used to optimize the bolus tim- ing. Set the ROI for monitoring scan in the abdominal aorta at the beginning of the scan range with trigger- ing threshold of 120 HU, or use manual triggering.

119