Dose Information

Effective mAs

In sequential scanning, the dose (Dseq) applied to the patient is the product of the tube current-time (mAs) and the CTDIw per mAs:

Dseq = DCTDIw x mAs

In spiral scanning, however, the applied dose (Dspiral) is influenced by the “classical“ mAs (mA x Rot Time) and in addition by the Pitch Factor. For example, if a Multislice CT scanner is used, the actual dose applied to the patient in spiral scanning will be decreased when the Pitch Factor is larger than 1, and increased when the Pitch Factor is smaller than 1. Therefore, the dose in spiral scanning has to be corrected by the Pitch Factor:

Dspiral = (DCTDIw x mA x Rot Time)/Pitch Factor

To make it easier for the users, the concept of the “effective“ mAs was introduced with the SOMATOM Multislice scanners.

The effective mAs takes into account the influence of pitch on both the image quality and dose:

Effective mAs = mAs/Pitch Factor

To calculate the dose you simply have to multiply the CTDIw per mAs with the effective mAs of the scan:

Dspiral = DCTDIw x effective mAs