Thorax
| ThorFast | 2nd reconstr. |
kV | 130 |
|
|
|
|
Effective mAs | 60 |
|
|
|
|
Rotation time | 0.8 sec |
|
|
|
|
Slice collimation | 5.0 mm |
|
|
|
|
Slice width | 8.0 mm | 8.0 mm |
|
|
|
Feed/Rotation | 20.0 mm |
|
|
|
|
Pitch Factor | 2.0 |
|
|
|
|
Increment | 8.0 mm | 8.0 mm |
|
|
|
Kernel | B41s | B70s |
|
|
|
CTDIVol | 6.47 mGy |
|
|
|
|
Effective dose | Male: |
|
| 3.38 mSv |
|
Female: 4.39 mSv
Contrast medium IV injection
Start delay | 25 – 30 sec. |
Flow rate | 2.5 ml/sec. |
|
|
Total amount | 80 ml |
Hints
•You could repeat the same protocol by simply click- ing the chronicle with the right mouse button for “repeat“, e.g., when both
•For lung cancer evaluation, this protocol can be com- bined with the protocol “Neck Routine”.
105