Head
Orbita
Indications:
Spiral mode for orbital studies, e.g., fracture.
A range of 2 cm will be covered in 12 sec.
| 
 | Orbita | 2nd reconstr. | 
| kV | 130 | 
 | 
| 
 | 
 | 
 | 
| Effective mAs | 100 | 
 | 
| 
 | 
 | 
 | 
| Rotation time | 1.0 sec | 
 | 
| 
 | 
 | 
 | 
| Slice collimation | 1.0 mm | 
 | 
| 
 | 
 | 
 | 
| Slice width | 3.0 mm | 1.25 mm | 
| 
 | 
 | 
 | 
| Feed/Rotation | 2.0 mm | 
 | 
| 
 | 
 | 
 | 
| Pitch Factor | 1.0 | 
 | 
| 
 | 
 | 
 | 
| Increment | 3.0 mm | 0.8 mm | 
| 
 | 
 | 
 | 
| Kernel | H70s | H70s | 
| 
 | 
 | 
 | 
| CTDIVol | 22.77 mGy | 
 | 
| 
 | 
 | 
 | 
| Effective dose | Male: | 
 | 
| 
 | 0.10 mSv | 
 | 
Female: 0.11 mSv
Contrast medium IV injection
| Start delay | 60 sec | 
| Flow rate | 2 ml/sec | 
| 
 | 
 | 
| Volume | 50 – 60 ml | 
84