Scan and Reconstruction

Pitch

In single slice CT:

Pitch = table movement per rotation/slice collimation

E.g.,: slice collimation = 5 mm,

table moves 5 mm per rotation, then pitch = 1.

With the Siemens Multislice CT, we differentiate between:

Feed/Rotation, the table movement per rotation

Volume Pitch, table movement per rotation/single slice collimation.

Pitch Factor, table movement per rotation/complete slice collimation.

E.g., slice collimation = 2 x 5 mm, table moves 10 mm per rotation, then Volume Pitch = 2, Pitch Factor = 1.

With the SOMATOM Spirit, the pitch, slice, collimation, rotation time, and scan range can be adjusted. The pitch factor can be selected from 0.5-2.

Pitch 1

Pitch 1.5

Pitch 2

Pitch Models