syngo Volume Evaluation

Restrictions for Linking and Unlinking

There are some restrictions for linking and unlinking:

Linking can be done on user-defined ROIs as well as on interpolated ROIs. If you link an interpolated ROI (dotted line), it becomes a user-confirmed ROI (solid line).

You cannot unlink interpolated ROIs. The unlinking only works with user-confirmed ROIs.

If you modify the volume (for example, move a ROI) before linking an ROI to the already selected ROI, the marking is reset.

Evaluation

Bounding voxels, where no voxel lies within the entered range of HU values, stops region growing. Check the depth of the volume in the coronal and sagittal view. To complete region growing inside the entire volume, set multiple seed points if necessary.

You can edit the thresholds even after the volume has been evaluated.

You can modify or delete the last evaluated volume.