Pelvis
| Hip | 2nd reconstr. |
kV | 130 |
|
|
|
|
Effective mAs | 70 |
|
|
|
|
Rotation time | 1.0 sec |
|
|
|
|
Slice collimation | 1.5 mm |
|
|
|
|
Slice width | 3.0 mm | 2.0 mm |
|
|
|
Feed/Rotation | 3.0 mm |
|
|
|
|
Pitch Factor | 1.0 |
|
|
|
|
Increment | 3.0 mm | 1.5 mm |
|
|
|
Kernel | B70s | B70s |
|
|
|
CTDIVol | 7.55 mGy |
|
|
|
|
Effective dose | Male: |
|
| 1.78 mSv |
|
| Female: |
|
| 1.66 mSv |
|
Hints
•In case of 3D study only, images should be recon- structed with at least 50% overlap in image recon- struction and kernel B20s.
•If only one side is examined, it is advisable to enter the side in the comment line on the Routine card.