Vascular

Contrast medium IV injection

Start delay

6 – 10 sec.

Flow rate

4 ml/sec.

 

 

Total amount

80 – 100 ml

Hints

CARE Bolus may be used to optimize the bolus tim- ing.

Set the ROI for monitoring scan in the pulmonary trunk with triggering threshold of 120 HU, or use manual triggering.

You could repeat the same protocol by simply click- ing the chronicle with the right mouse button for "repeat".

171