Head
| BaseSeq | CerebrumSeq | |
|
|
|
|
kV | 130 | 130 | |
|
|
|
|
Effective mAs | 240 | 240 | |
|
|
|
|
Rotation time | 1.5 sec | 1.5 sec | |
|
|
|
|
Slice collimation | 1.5 mm | 4.0 mm | |
|
|
|
|
Slice width | 3.0 mm | 8.0 mm | |
|
|
|
|
Feed/Scan | 3.0 mm | 8.5 mm | |
|
|
|
|
Kernel | H31s | H31s | |
|
|
|
|
CTDIVol | 54.65 mGy | 51.43 mGy | |
|
|
|
|
Effective dose | Male: | Male: 1.52 mSv | |
| 0.77 mSv | Female: | |
| Female: | 1.68 mSv | |
| 0.80 mSv |
|
|
|
|
| |
|
|
| |
Contrast medium IV injection |
|
| |
Start delay | 60 sec. |
|
|
|
|
|
|
Flow rate | 2 ml/sec. |
|
|
|
|
|
|
Volume | 50 – 60 ml |
|
|
Hints
•An automatic bone correction allows for improved head image quality, without any additional
•In order to optimize image quality versus radiation dose, scans are provided within a maximum scan field of 300 mm with respect to the