Children

Abdomen_Child

Indications:

Spiral mode for routine studies in the region of abdo- men and pelvis, e.g., tumors, lymphoma, abscesses, post-traumatic changes, etc.

A range of 30 cm will be covered in 32 sec.

 

Abdomen

 

 

kV

130

 

 

Effective mAs

41

 

 

Rotation time

1.0 sec

 

 

Slice collimation

2.5 mm

 

 

Slice width

5.0 mm

 

 

Feed/Rotation

10.0 mm

 

 

Pitch Factor

2.0

 

 

Increment

5.0 mm

 

 

Kernel

B41s

 

 

CTDIVol

4.42 mGy

 

 

Effective dose

Male: 4.83 mSv*

 

Female: 7.40 mSv*

 

 

*The conversion factor for a 7-year-old child, and a scan range of 300 mm was used.

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/

 

Access site

 

 

Total amount

1 – 2 ml per kg of body weight