Children
| 
 | ThorRoutine | 2nd | 
| 
 | 
 | reconstruction | 
| kV | 130 | 
 | 
| 
 | 
 | 
 | 
| Effective mAs | 26 | 
 | 
| 
 | 
 | 
 | 
| Rotation time | 1.0 sec | 
 | 
| 
 | 
 | 
 | 
| Slice | 2.5 mm | 
 | 
| collimation | 
 | 
 | 
| 
 | 
 | 
 | 
| Slice width | 5.0 mm | 5.0 mm | 
| 
 | 
 | 
 | 
| Feed/Rotation | 10.0 mm | 
 | 
| 
 | 
 | 
 | 
| Pitch Factor | 2.0 | 
 | 
| 
 | 
 | 
 | 
| Increment | 5.0 mm | 5.0 mm | 
| 
 | 
 | 
 | 
| Kernel | B41s | B60s | 
| 
 | 
 | 
 | 
| CTDIVol | 2.80 mGy | 
 | 
| 
 | 
 | 
 | 
| Effective dose | Male: 1.46 mSv* | 
 | 
| 
 | Female: 1.72 mSv* | |
| 
 | 
 | 
 | 
*The conversion factor for a 
Contrast medium IV injection
| Start delay | exam dependent | 
| Flow rate | dependent upon needle size/ | 
| 
 | Access site | 
| 
 | 
 | 
| Total amount | 1 – 2 ml per kg of body weight | 
Hints
•Children with a body weight of more than 55 kg should be examined with an adult protocol.
•The first and second recon jobs are defined for visu- alization of the mediastinum and the lungs, respec- tively.
•Please change the mAs value according to the weight of the child:
45 – 54 kg 37 mAs.