Lower Extremities

FootHR

Indications:

Spiral mode for high resolution studies of the foot, e.g., masses, trauma, disorders of the joint etc.

A range of 10 cm will be covered in 52 sec.

 

FootHR

2nd reconstruction

kV

130

 

 

 

 

Effective mAs

30

 

 

 

 

Rotation time

1.0 sec

 

 

 

 

Slice collimation

1.0 mm

 

 

 

 

Slice width

3.0 mm

1.25 mm

 

 

 

Feed/Rotation

2.0 mm

 

 

 

 

Pitch Factor

1.0

 

 

 

 

Increment

3.0 mm

0.8 mm

 

 

 

Kernel

U90s

U90s

 

 

 

CTDIVol

3.23 mGy

 

 

 

Effective dose

Male: 0.01 mSv

 

Female: 0.01 mSv

Hint

For image reconstruction of soft tissue, use kernel B31s.