Siemens Software version syngo CT 2005C manual Dose Information

Models: Software version syngo CT 2005C

1 321
Download 321 pages 19.7 Kb
Page 31
Image 31

Dose Information

For spiral scan protocols, the indicated mAs is the effective mAs per image. The correlation between tube current mA and effective mAs of spiral scans on a Mul- tislice CT scanner is given by the following formula:

Effective mAs = mA x RotTime/Pitch Factor

Pitch Factor =

Feed/Rot

nrow x Slice collimation

mA = effective mAs x Pitch Factor RotTime

where Slice collimation refers to the collimation of one detector row, and nrow is the number of used detector rows.

31

Page 31
Image 31
Siemens Software version syngo CT 2005C manual Dose Information