Head

InnerEarHRSeq

Indications:

Sequential mode for inner ear studies, e.g., inflamma- tory changes, tumorous processes of pyramids, cere- bellopontine angle tumors, post-traumatic changes, etc.

A typical gantry tilt of -20 degree is predefined. A range of 2.5 cm will be covered.

 

InnerEarSeq

 

 

kV

130

 

 

Effective mAs

140

 

 

Rotation time

1.5 sec

 

 

Slice collimation

1.0 mm

 

 

Slice width

1.0 mm

 

 

Feed/Scan

2.0 mm

 

 

Kernel

H80s

 

 

CTDIVol

31.88 mGy

 

 

Effective dose

Male: 0.32 mSv

 

Female: 0.38 mSv

 

 

Contrast medium IV injection

 

 

Start delay

60 sec.

 

 

Flow rate

2 ml/sec.

 

 

Total amount

50 – 60 ml

 

 

78

Page 78
Image 78
Siemens Software version syngo CT 2005C manual InnerEarHRSeq, InnerEarSeq, 130