Head
InnerEarHRSeq
Indications:
Sequential mode for inner ear studies, e.g., inflamma- tory changes, tumorous processes of pyramids, cere- bellopontine angle tumors,
A typical gantry tilt of
| InnerEarSeq |
|
|
kV | 130 |
|
|
Effective mAs | 140 |
|
|
Rotation time | 1.5 sec |
|
|
Slice collimation | 1.0 mm |
|
|
Slice width | 1.0 mm |
|
|
Feed/Scan | 2.0 mm |
|
|
Kernel | H80s |
|
|
CTDIVol | 31.88 mGy |
|
|
Effective dose | Male: 0.32 mSv |
| Female: 0.38 mSv |
| |
| |
Contrast medium IV injection | |
|
|
Start delay | 60 sec. |
|
|
Flow rate | 2 ml/sec. |
|
|
Total amount | 50 – 60 ml |
|
|
78