Specials

PolyTrauma

Two times 2 combined recon jobs are predefined, head with neck and thorax with abdomen.

A range for the Head of 11 cm will be covered in

20 sec., a range for the Neck of 14 cm will be covered in 30 sec.

 

Head

Neck

kV

130

130

 

 

 

Effective mAs

110

90

 

 

 

Rotation time

1.5 sec

1.0 sec

 

 

 

Slice collimation

5.0 mm

2.5 mm

 

 

 

Slice width

6.0 mm

3.0 mm

 

 

 

Feed/Rotation

10.0 mm

5.0 mm

 

 

 

Pitch Factor

1.0

1.0

 

 

 

Increment

6.0 mm

3.0 mm

 

 

 

Kernel

H31s

B50s

 

 

 

CTDIVol

25.05 mGy

9.70 mGy

 

 

 

Effective dose

Male:

Male:

 

1.02 mSv

1.69 mSv

 

Female:

Female:

 

1.08 mSv

1.81 mSv

 

 

 

Take a new Topogram for the thorax and abdomen ranges.