Children

BodyAngio08s

Indications:

Spiral mode for abdominal CT Angio studies, e.g., vas- cular abnormalities, aneurysms, etc.

A range of 10 cm will be covered in 15 sec.

 

BodyAngio

2nd reconstr.

kV

80

 

 

 

 

Effective mAs

35

 

 

 

 

Rotation time

0.8 sec

 

 

 

 

Slice

1.5 mm

 

collimation

 

 

 

 

 

Slice width

3.0 mm

2.0 mm

 

 

 

Feed/Rotation

6.0 mm

 

 

 

 

Pitch Factor

2.0

 

 

 

 

Increment

3.0 mm

1.5 mm

 

 

 

Kernel

B41s

B41s

 

 

 

CTDIVol

0.96 mGy

 

 

 

 

Effective dose

Male:

 

0.62mSv*

Female:

0.72mSv*

* The conversion factor for a 7-year-old child, and a scan range of 100 mm was used.

236

Page 236
Image 236
Siemens Software version syngo CT 2005C manual BodyAngio08s