STATus Subsystem

STATus:OPERation:VXITrigger:CONDition?

Queries the contents of the VXI TTL Trigger Monitor Condition register. A value of one in a bit indicates the corresponding condition bit has transitioned from high to low. Unused bits always return a value of zero. The definition of the bits in this register can be found in STATus:OPERation:VXITrigger command.

The enhanced monitor latches the state of VXI TTL Trigger lines into a hardware register. This register cleared upon VXI Monitor initialization. This query returns the state of this register.

There is no command associated with this query.

Syntax

ST￿Tus￿OPERation￿VXITrigger￿CONDition￿

Parameters

 

 

 

Parameter

 

Query response

 

Not applicable

<NR1>

Reset Value

 

 

Not Applicable

 

Errors and Events

None

 

 

Dependencies

None

 

 

Examples

Query:

ST￿T￿OPER￿VXIT￿COND￿

 

 

Response:

0

 

Related Commands STATus:OPERation:VXITrigger?

STATus:OPERation:VXITrigger:ENABle

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual