Adjustment Procedures
SYSTem:SERial
SYSTem:SERial?
This command stores the serial number of the instrument in nonvolatile RAM.
The query returns the serial number previously stored.
The command converts all lower case characters in the serial number to upper case.
Syntax | SYSTem:SERial <Serial Number> | ||
| SYSTem:SERial? | ||
Parameters |
|
|
|
<Parameter> |
| Query response | |
| <string> |
| <string> |
| The serial number must start with an alphabet- |
| |
| ical character followed by an optional |
| |
| alphabetical character and by a maximum of |
| |
| six digits. |
|
|
Reset Value |
|
| |
Not Applicable | |||
Errors and Events | ±221, ªSettings Conflict.º | ||
| You attempted to execute a command that is protected by a password. See the | ||
| SYSTem:PASSword command beginning on page 7±5 for more information. | ||
| ±220, ªParameter Errorº | ||
| Invalid serial number. | ||
Dependencies | None |
|
|
Examples | Command: SYST:SER "B123456" | ||
| Query: | SYST:SER? | |
| Response: | •B123456" | |
Related Commands | *IDN |
|
|
7±8 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |