SYSTem Subsystem
SYSTem:SYSReset
SYSTem:SYSReset?
Sets or queries whether the VXI SYSRESET signal can reset the Enhanced
Monitor.
When SYSRESET is enabled, asserting the VXI SYSRESET signal will reset the Enhanced Monitor. When disabled, asserting the VXI SYSRESET signal only initializes the Enhanced Monitor VXI Interface; all other Enhanced Monitor functions are not affected.
The factory default setting is disabled (0).
NOTE. If SYSRESET is enabled and the mainframe is powered down, communication through the
Syntax | SYSTem:SYSReset <Boolean> | |||
| SYSTem:SYSReset? | |||
Parameters |
|
|
|
|
| <Boolean> |
| Query response | |
|
| <NRf> |
| <NR1> |
|
| N ≠ 0 |
| 1 |
| 0 |
| 0 | |
|
| ON |
| 1 |
|
| OFF |
| 0 |
Reset Value |
|
|
| |
| Not Applicable | |||
Errors and Events |
| ±224, ªIllegal Parameter Valueº | ||
|
| You attempted to set the parameter to an illegal value. | ||
Dependencies |
| None |
|
|
Examples |
| Command: SYST:SYSR ON | ||
| Query: | SYST:SYSR? | ||
|
| Response: | 1 |
|
Related Commands | SYSTem:SYSReset:IMMediate |