Adjustment Procedures
9.Refer to Table 7±1 and set up the PC
Table 7±1: PC
Procedure
Setting | |
|
|
Baud Rate | 9600 |
|
|
Data Bits | 8 |
|
|
Stop Bits | 1 |
|
|
Parity | None |
|
|
Flow Control | Xon/Xoff |
|
|
10.Make sure that the mainframe has had at least five minutes to warm up before continuing this procedure.
11.Send the following commands to enable the password and to calibrate the IntelliFrame Mainframe:
SYST:PASS •VX1410" CAL?
12.Check that the CAL? query returns a value of zero to indicate no error has been detected during the calibration.
NOTE. Steps 13 and 14 are only necessary if the you have replaced the Enhanced
Monitor. If you haven't replaced the Enhanced Monitor go to step 15.
13.Note the serial number on the rear panel of the mainframe.
14.Send the following command to set up the serial number in the Enhanced Monitor:
SYST:SER •Bxxxxxx"
The value of Bxxxxxx is the serial number as noted from the rear of the mainframe.
7±2 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |