Adjustment Procedures

9.Refer to Table 7±1 and set up the PC RS-232 port configurations. The RS-232 settings must match those of the mainframe.

Table 7±1: PC RS-232 settings for the calibration

Procedure

RS-232 Parameter

Setting

 

 

Baud Rate

9600

 

 

Data Bits

8

 

 

Stop Bits

1

 

 

Parity

None

 

 

Flow Control

Xon/Xoff

 

 

10.Make sure that the mainframe has had at least five minutes to warm up before continuing this procedure.

11.Send the following commands to enable the password and to calibrate the IntelliFrame Mainframe:

SYST:PASS •VX1410" CAL?

12.Check that the CAL? query returns a value of zero to indicate no error has been detected during the calibration.

NOTE. Steps 13 and 14 are only necessary if the you have replaced the Enhanced

Monitor. If you haven't replaced the Enhanced Monitor go to step 15.

13.Note the serial number on the rear panel of the mainframe.

14.Send the following command to set up the serial number in the Enhanced Monitor:

SYST:SER •Bxxxxxx"

The value of Bxxxxxx is the serial number as noted from the rear of the mainframe.