SYSTem Subsystem

SYSTem:ERRor?

This query returns the next entry from the Error/Event queue. This is an alias command with the STATus:QUEue[:NEXT]? command. There is no associated command for this query.

The Error/Event queue is cleared at power up, upon receipt of *CLS, and upon reading the last item from the queue.

Syntax SYSTem:ERRor?

Parameters

Parameter

 

Query response

 

Not Applicable

<NR1>, <string>

Reset Value

 

 

Not Applicable

 

Errors and Events

±350, ªQueue Overflowº

 

 

The Error/Event queue overflowed due to the execution errors of other com-

 

mands. The SYST:ERR? query, itself, cannot generate an error.

Dependencies

Reading the event clears the entry from the queue.

Examples

Query:

SYST:ERR?

 

 

Response:

0, •No error"

 

Related Commands

STATus:PRESet

 

 

STATus:QUEue?

 

 

STATus:QUEue:ENABle

 

 

*CLS

 

 

3±124

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual