Syntax

NOTE. Do not include the <, >, {,}, or symbols when entering the actual value for a parameter.

Separate multiple parameters after a command header with a comma. For example:

STATus:QUEStionable:TEMPerature:LIMit DELTa1,30

Table 3±13 defines the command and response parameter types for the Intelli- Frame Mainframe.

Table 3±13: Parameter types for syntax descriptions

Parameter type

Description

Example

 

 

 

Boolean

Boolean number or value (command parameter,

ON or 1

 

response returns a 1 or 0)

OFF or 0

 

 

 

Block

A specified length of block-formatted data

#212xxxxxxxxxxxx

 

(command parameter)

where 2 indicates that

 

 

the following 2 digits

 

 

(12) specify the length

 

 

of the following block

 

 

data (xxx...)

 

 

 

Nondecimal numeric

Integers in hexadecimal (#H), octal (#Q), or

#H00FF, #B010101,

 

binary (#B) format (command parameter).

#Q0753

 

 

 

NR1

Integers (response parameter)

0, 1, 15

 

 

 

NR2

Decimal numbers (response parameter)

1.2, 3.141516, 6.5

 

 

 

NRf

Integer, decimal, or floating point number as well

1, 3E+1, MAX, 2.6

 

as MAX and MIN for setting the largest or

 

 

smallest value the instrument allows. (command

 

 

parameter)

 

 

 

 

discrete

A list of specific values (command or response

P24, P5EXT, N2, ...

 

parameter)

 

 

 

 

String

Alphanumeric characters within quotation marks

ªNo errorº

 

(command or response parameter)

 

 

 

 

Parameter values that appear in this manual are often enclosed in curly braces ({}) separated by a vertical line. This vertical line means the same thing as the word ªor.º For example, values for the parameter <Boolean> are

{01OFFON}

This is the same thing as saying ª0 or 1 or OFF or ON.º Any single value is a valid parameter.

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual

3±23