Adjustment Procedures

15.VX1410A with Option 1M or a VX1420A: Use then the front panel Next

and Prev buttons to scroll through the delta temperature of the slots. All delta temperatures should read 0 °C ±1 °C.

16.Power down the mainframe.

17.This completes the calibration procedure.

The rest of this section describes the CALibration and SYSTem:PASSword commands in detail.

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual

7±3