Adjustment Procedures
15.VX1410A with Option 1M or a VX1420A: Use then the front panel Next
and Prev buttons to scroll through the delta temperature of the slots. All delta temperatures should read 0 °C ±1 °C.
16.Power down the mainframe.
17.This completes the calibration procedure.
The rest of this section describes the CALibration and SYSTem:PASSword commands in detail.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 7±3 |