STATus Subsystem
STATus:OPERation?
Returns the contents of the Operation Event register. A value of one in a bit indicates the corresponding condition bit changed from false to true. Unused bits always return a value of zero. The register is cleared by the *CLS command.
There is no command associated with this query.
Syntax | STATus:OPERation[:EVENt]? |
| |
Parameters |
|
|
|
| Parameter | Query response | |
|
| Not applicable | <NR1> |
|
|
|
|
|
|
|
|
|
| Operation status word bits | Definition |
|
|
|
|
|
| D0 | CALibrating |
|
|
|
|
|
| D1 | Not Used |
|
|
|
|
|
| D2 | Not Used |
|
|
|
|
|
| D3 | Not Used |
|
|
|
|
|
| D4 | Not Used |
|
|
|
|
|
| D5 | VXI Triggers |
|
|
|
|
|
| D6 | Not Used |
|
|
|
|
|
| D7 | Not Used |
|
|
|
|
|
| D8 | VXI Monitor |
|
|
|
|
|
| D9 | Mainframe Power Down |
|
|
|
|
|
| D10 | VXI Interrupts |
|
|
|
|
|
| D11 | Not Used |
|
|
|
|
|
| D12 | Not Used |
|
|
|
|
|
| D13 | Not Used |
|
|
|
|
|
| D14 | Not Used |
|
|
|
|
|
| D15 | Not Used |
Reset Value |
|
|
|
0 |
| ||
Errors and Events |
| None |
|
Dependencies |
| Reading the register clears it. |
|
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 3±33 |