Specifications
This chapter contains the complete specifications for the IntelliFrame Main- frame. Within each section, the specifications are arranged in functional groups such as: AC Power Source, Secondary Power, Cooling, Certifications and compliances, Environmental, Backplane, Display System and Mechanical.
All specifications are warranted unless unless labeled typical. Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the n symbol are checked in the Performance Verification chapter beginning on page 5±1.
NOTE. All VX1410A specifications listed in this chapter also apply to VX1411A mainframes that have been upgraded. All VX1420A specifications also apply to upgraded VX1410A and VX1411A mainframes.
Table 4±1: AC Power Source
Characteristic | Description | |
|
|
|
Source Voltage | 90 VRMS to 250 VRMS, continuous range | |
Source Frequency | 115 V Operation: 45 to 66 Hz or 360 to 440 Hz | |
|
| 230 V Operation: 45 to 66 Hz |
|
|
|
Power Consumption | VX1410A:1350 W line power maximum | |
|
| VX1420A:1450 W line power maximum |
|
|
|
Fuse Rating |
| |
| 90 V ± 132 V Operation | 0.25 in ⋅1.25 in, Slow Blow, 20 A, 250 V |
|
|
|
| 103 V ± 250 V Operation | 0.25 in ⋅1.25 in, Fast Blow, 15 A, 250 V |
|
|
|
| 207 V ± 250 V Operation | 5 mm ⋅20 mm, Fast Blow, 6.3 A, 250 V |
|
| |
Inrush Surge Current | 70 A maximum | |
|
| |
Input Current | VX1410A:15 A maximum at 90 VRMS | |
|
| 6.3 A maximum at 207 VRMS |
|
| VX1420A:16.5 A maximum at 90 VRMS |
|
| 6.3 A maximum at 207 VRMS |
Power Factor Correction | 0.99 @ 60 Hz operation | |
|
| 0.95 @ 400 Hz operation |
|
| |
Power Disconnect | Front Panel On/Standby. No primary switch on rear | |
|
| panel. Power cord provides main power disconnect. The |
|
| Front Panel On/Standby switch may be disabled for |
|
| remote operation. Enhanced monitor provides ability to |
|
| turn the instrument on/off under program control. |
|
|
|
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 4±1 |