Specifications

This chapter contains the complete specifications for the IntelliFrame Main- frame. Within each section, the specifications are arranged in functional groups such as: AC Power Source, Secondary Power, Cooling, Certifications and compliances, Environmental, Backplane, Display System and Mechanical.

All specifications are warranted unless unless labeled typical. Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the n symbol are checked in the Performance Verification chapter beginning on page 5±1.

NOTE. All VX1410A specifications listed in this chapter also apply to VX1411A mainframes that have been upgraded. All VX1420A specifications also apply to upgraded VX1410A and VX1411A mainframes.

Table 4±1: AC Power Source

Characteristic

Description

 

 

 

Source Voltage

90 VRMS to 250 VRMS, continuous range

Source Frequency

115 V Operation: 45 to 66 Hz or 360 to 440 Hz

 

 

230 V Operation: 45 to 66 Hz

 

 

 

Power Consumption

VX1410A:1350 W line power maximum

 

 

VX1420A:1450 W line power maximum

 

 

 

Fuse Rating

 

 

90 V ± 132 V Operation

0.25 in 1.25 in, Slow Blow, 20 A, 250 V

 

 

 

 

103 V ± 250 V Operation

0.25 in 1.25 in, Fast Blow, 15 A, 250 V

 

 

 

 

207 V ± 250 V Operation

5 mm 20 mm, Fast Blow, 6.3 A, 250 V

 

 

Inrush Surge Current

70 A maximum

 

 

Input Current

VX1410A:15 A maximum at 90 VRMS

 

 

6.3 A maximum at 207 VRMS

 

 

VX1420A:16.5 A maximum at 90 VRMS

 

 

6.3 A maximum at 207 VRMS

Power Factor Correction

0.99 @ 60 Hz operation

 

 

0.95 @ 400 Hz operation

 

 

Power Disconnect

Front Panel On/Standby. No primary switch on rear

 

 

panel. Power cord provides main power disconnect. The

 

 

Front Panel On/Standby switch may be disabled for

 

 

remote operation. Enhanced monitor provides ability to

 

 

turn the instrument on/off under program control.

 

 

 

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual